NOVUS SMARTMETER User Manual

Page 3

Advertising
background image

SmartMeter

NOVUS PRODUTOS ELETRÔNICOS LTDA.

3/4

CONFIGURATION CYCLE

To enter the configuration cycle, press

BACK

and

SET

. The meter

displays

, the first parameter of the cycle. To access the

remaining parameters ofthis cycle, press

SET

. Keys

and

make possible to change the parameters defined. To access the

measures screen, go to the end of this cycle or press keys

BACK

and

SET

again.

The parameters of the configuration cycle are:

Selects the channel 1 sensor according to Table 1. Channel 1 sensor

is chosen regardless of the channel 2 sensor selection.

Selects the channel 2 sensor according to Table 1. Channel 2 sensor

is chosen regardless of the channel 1 sensor selection. Channel 2

indication can be disabled.

Code

Input

Channel 2 - disabled

0-50 mV

Pt100

B thermocouple

S thermocouple

R thermocouple

N thermocouple

E thermocouple

T thermocouple

K thermocouple

J thermocouple

Table 1

(Decimal Point Position)

Enables the display of decimal points for channel 1 measures.

Parameter not valid for thermocouples

R, S and B. For the 0-50 mV

input, defines the decimal point position.

(Decimal Point Position)

Enables the display of decimal points for channel 2 measures.

Parameter not valid for thermocouples

R, S and B. For the 0-50 mV

input, defines the decimal point position.

(Lowest indication limit of channel 1 for 0-50 mV)

Determines the inferior limit of the user-created meter indication

range, when the 0-50 mV input signal is selected. Parameter not

displayed for the other input types.

(Highest indication limit of channel 1 for 0-50 mV)

Determines the upper limit of the user-created meter indication range,

when the 0-50 mV input signal is selected. Parameter not displayed

for the other input types.

(Lowest indication limit of channel 2 for 0-50 mV)

Determines the inferior limit of the user-created meter indication

range, when the 0-50 mV input signal is selected. Parameter not

displayed for the other input types.

(Highest indication limit of channel 2 for 0-50 mV)

Determines the upper limit of the user-created meter indication range,

when the 0-50 mV input signal is selected. Parameter not displayed

for the other input types.

(Indication Offset, channel 1)

Allows for an offset in the temperature value displayed in channel 1.

The offset value is simply added to the value measured by the

sensor. The offset can be of 10 % of the selected sensor maximum

range, upwards or downwards.

(Indication Offset, channel 2)

Allows for an offset in the temperature value displayed in channel 2.

The offset value is simply added to the value measured by the

sensor. The offset can be of 10 % of the selected sensor maximum

range, upwards or downwards.

(Beep – Sound alarm)

It enables or disables the sound alarm

: disabled
: enabled

Note: The

ALARM

message is displayed regardless of the sound

alarm status (on or off). To stop beeping while an alarm condition is

present, press any key.

(shut-off – Auto power off)

It makes possible to define a time interval in which the meter powers

off automatically, provided no key is pressed within the interval.
Adjustable between 3 and 120 minutes.

!

!

!

!

(Unit – Temperature indication unit)

It defines the temperature unit of the meter.

ºC for degrees Celsius and ºF Fahrenheit.

!

!

!

!

(Noise filter)

It adapts the meter frequency to the mains voltage in order to

optimize the internal noise filter. Such adaptation reduces possible

temperature oscillations, resulting from the mains frequency. Options

are:

50 - for 50 Hz mains.
60 - for 60 Hz mains.

(Special functions)

Defines whether special functions HOLD, AVG, MAX, MIN and

RELAT will be applied to channels 1 and 2 or only to channel 1.

Ch1 - special functions applied only to channel 1.
Ch1 - special functions applied to channels 1 and 2

4.4 CALIBRATION CYCLE
To enter the calibration cycle, press

BACK

and

SET

for 10 seconds. The

meter displays

, the first parameter of the cycle. To access the

remaining parameters of this cycle, press

SET

.

Keys

and

make possible to change the parameters defined. To access the

measures screen, go to the end of this cycle or press keys

BACK

and

SET

again. The parameters of the configuration cycle are:

(Software version)

About information. Informs the software review installed.

""""

""""

""""

""""

(Serial Number)

About information. Informs the meter serial number. The four larger

numbers are the most significant and the smaller the less significant.

#$ $

#$ $

#$ $

#$ $

(

L

ow

C

alibration Channel 1)

INPUT OFFSET CALIBRATION: Adjusts the lowest limit of the input

signal indication range (offset). In order to change the indicated

value, it may be required to press

or

several times.

$ $

$ $

$ $

$ $

(

High

C

alibration Channel 1)

INPUT SPAN CALIBRATION: Adjusts the highest limit of the input

signal indication range (span). In order to change the indicated value,

it may be required to press

or

several times.

$ #

$ #

$ #

$ #

Adjusts the room temperature where the meter is held. Important

information for calculations and meter internal compensation.