Figure 02 – TEMP model dimensions

Ambient temperature channel: Internal thermistor.

Built-in cold junction compensation for thermocouples and wire resistance for Pt100.

Total accuracy:

Thermocouple K: ± 4.5°C (when maximum resoluti on is selected);

Pt100: ± 1.5°C (when maximum resolution is sel ected);

Ambient Temperature Channel: ± 1.5°C (20 minut es after connection to the USB port)

Note: It is possible to reduce (and even cut out) errors if after calibration the calibration points are
transferred to the device using the custom calibration screen (via software).

A/D resolution: Configurable from 15 to 11 bits.

Sampling rate: 8 to 128 samples per second, depending on resolution and number of channels enabled.

RHT Model (temperature and relative humidity)

Maximum measurement range: -20 to 100ºC and 0 to 100% RH (available also as dew point). See Figure 04

“Humidity sensor operating conditions”.

Dimensions: See Figure 03.

Figure 03 - RHT model dimensions

Accuracy: See Figure 04.

Note: It is possible to reduce (and even cut out) errors if after calibration the calibration points are
transferred to the device using the custom calibration screen (via software).