Figure 04 – Operating accuracy and conditions of RHT model

Sampling rate: one sample at every 1 or 3 seconds, depending on the resolution selected.

PC interface: Plug and Play USB (compatible with V1.1 and V2.0), virtual serial port interface, 1.5 m long cable.

Power supply: From the USB bus. Typical current: 30 mA

Supports custom multipoint calibration, which allows for an increase of system accuracy.

Handle operating environment: 0 to 60°C and 0 to 95% RH non-condensing.

Electromagnetic compatibility: EN 50081-2, EN 50082-2.

ABS case.

TESTS

This section shows the step response of some myPCProbe models to humidity (RHT model) and temperature (TEMP
model).

RHT MODEL

The myPCProbe-RHT was taken from an extremely dry environment and placed in an extremely wet environment.

Response time (0 to 90%) = 07 min 45 s

Figure 05 - RHT model humidity step response