NOVUS myPCProbe User Manual
Page 4
Advertising
myPCProbe
NOVUS PRODUTOS ELETRONICOS LTDA www.novusautomation.com
4/11
TEMP MODEL
In these tests, a calibration furnace was used to provide the required final temperature. Basal temperature was room
temperature (around 25ºC; 77 ºF).
150 x 3 mm probe, thermocouple K
Features:
See model number 2 in Table 01.
Probe inserted 130 mm in the furnace.
Response time (0 to 90%) = 2 min 12 s
Figure 06 – TEMP model temperature step response with 150 x 3 mm, thermocouple K probe
300 x 6 mm probe, Pt100
Features:
See probe number 5 in Table 01.
Probe inserted 150 mm in the furnace.
Response time (0 to 90%) = 3 min 46 s
Figure 07 – TEMP model temperature step response with 300 x 6 mm, Pt100 probe