Easycal, User guide – BrandTech Scientific EASYCAL 4.0 User Manual
Page 10
10
EASYCAL
™
User Guide Version 4.0 · Status: June 2008
EASYCAL
™
User Guide
Note:
The primary data according to GLP instructions will be transmitted without conversion with Factor “Z”.
Double-check all entries in the window, and then click on the button “Entry of tolerance limits” which will
take you to the next screen.
☞ Click on “Cancel” to close the current measuring result entry session. A question will pop up, whether you
want to return to the main menu.
3.1.2 Entry of error limits
On the “Tolerance limits” screen you determine the error limits applicable to your area of work.
These values will be used later on for the analysis of all your test results.
☞ Click on “Data analysis” to display the results of the entry of error limits for the test. Instruments and values
saved at this stage will appear among “Previous test records” and cannot subsequently be altered (excep-
tion: “Note” field).
☞ Click on “Back” to return to the entry of measuring results.
☞ Click on “Close” to close the current operation. A question will pop up, wether you want to return to the
main menu.
The test was carried out according to:
Click on the arrow to the right of the entry field to display a list of the 4 different available standards.
Error limits from Operating Manual:
the default values can be edited.
Error limits acc. to ISO 8655 (de-
fault setting)
Define your own error limits in %.
A = Accuracy of measurements
CV = Coefficient of variation of
measured
values