Easycal, User guide – BrandTech Scientific EASYCAL 4.0 User Manual
Page 14
14
EASYCAL
™
User Guide Version 4.0 · Status: June 2008
EASYCAL
™
User Guide
Entry of measuring results
☞ The entry fields for measuring results have a
white background.
☞ In the first white column, click into the first entry
field “x1”.
☞ Manual entry of results:
Enter measured value x1 and confirm with “EN-
TER”. This takes you automatically to the next
entry field. Enter the remaining results.
Note
Use this entry field to record in the test record any damages or functional defects of the tested instrument.
Test Volume/Nominal Values
The right-hand section of the entry window contains the table for entering the values you have found in the gravi-
metric test. In the “Test Volume” line in the table header, the following volumes are suggested for the selected
instruments:
Column 1:
Nominal volume
Column 2:
50 % of nominal volume
Column 3:
10 % of nominal volume
The test volumes can be modified as required by overwriting them.
Instrument type
Number required
Calibrated to contain (In)
1
Calibrated to deliver (Ex)
3
For batch testing, up to 30 measured values are allowed.
Note:
The primary data according to GLP instructions will be transmitted without conversion according to
Factor “Z”. Double-check all entries in the window, and then click on the button “Entry of tolerance limits“,
which will take you to the next screen.
Number of possible measured values for individual testing
☞ Automatic transfer of results from the balance (optional):
If the balance is connected, results are transferred automatically from the balance (Professional Version). To
transfer the weighed value, press either the “Print” button on the balance, or the F2 key on your PC keyboard.
Press “ENTER” to confirm the weighed value. This takes you automatically to the next entry field (see Section
5.5 ”Connecting a balance”).
☞ If other columns have a white background, these have to be filled in with results also.