Easycal, User guide – BrandTech Scientific EASYCAL 4.0 User Manual
Page 15
EASYCAL
™
User Guide Version 4.0 · Status: June 2008
15
EASYCAL
™
User Guide
3.2.2 Entry of error limits
On the “Entry of tolerance limits” screen you determine the error limits applicable to your area of work. These
values will be used later on for the analysis of all your test results.
☞ Click on the “Data analysis” button to display the results of the Entry of error limits for the test.
Instruments and values saved at this stage will appear in the Test History.
☞ Click on the “Back” button to return to “Entry of measuring results”.
☞ Click on the “Close” button to end the current operation. You were asked if you want to return to the
main menu.
3.2.3 “Data analysis”
The “Data analysis” screen can only be opened if all values required for the analysis have been entered.
You cannot make any entries in this window.
Printer Setup
Modify the default values if required, and click “OK” to start printing. The program compiles the record from a
number of different components and sends this information to the selected printer. This may take some time,
depending on the PC you are using.
Declare instrument defect
If the check reports a fault in the device,
this field can be selected.
The device should then be checked for
leaks or defective components; if this
cannot be done in-house, please send it
in for repair.
User-specific error
limits
Error limits to DIN
Error limits stated by
manufacturer
Cancel the current operation. You can
decide whether you want to save the
entries you have made.
Back to “Entry of toler-
ance limits” screen
Start printing
the test report
Preview the test
report