EASYCAL

User Guide Version 4.0 · Status: June 2008

EASYCAL

User Guide

3.2.4 Save present test (permanently) and Intervall entry

4.1 Selecting records from history

When the “Previous test records” function is acti-
vated, the following screen is displayed:

Note:
When entering data for new instruments, please double-check the accuracy of your entries. Data saved in the
instrument table can no longer be modified.

4. Test History

4.2 Deleting records from history

Highlight the record to be deleted by moving the blue bar on this record (as illustrated above). Use the up/down
arrow keys

↑↓ or the mouse. Click on “Delete” to remove the data record from history. If you wish to delete several

records at a time, select several subsequent records. For this purpose, please press shift and click the left mouse
button to indicate the beginning or the end of the selection. Enter the password “EASYADMIN”. The record high-
lighted in blue will be deleted. For changing the password, see 5.4 “Manage EASYCAL

data”.

The “Data analysis” screen can only be opened if all
values required for the analysis have been entered.
When you close the “Data analysis” screen, you have
the option of saving the complete data record perma-
nently. A data record thus saved can be viewed later
under Previous test records”. Once the record has
been saved, no further changes can be made. At this
point you will also be asked at which interval the instru-
ment has to be tested again.Please enter the required
interval in months (1-99).
When the evaluation is complete, any newly entered
test operators and instruments are stored in the
relevant lists.

The “Previous test records” function allows you to view a history of earlier tests. All data records stored in the