Easycal, User guide – BrandTech Scientific EASYCAL 4.0 User Manual

Page 28

Advertising
background image

28

EASYCAL

User Guide Version 4.0 · Status: June 2008

EASYCAL

User Guide

Definitions in Test Record for Glass/Plastics

Deviation =
Maximum deviation from nominal value
in the sample

Range = |EL ¼| The measuring
uncertainty is acceptable if the “Range”
is 1/4 of the error limit of the volumetric
instrument.

Note:
The relative humidity during the test should be between 50% ± 15%. Enter it to the record.

Range (of measured values) =
highest value minus smallest value

Error limit (mean value). Statistical
value, since only part of the batch was
tested.