Crd5490-z – Cirrus Logic CRD5490 User Manual
Page 19
Advertising
CRD5490-Z
DS988RD1
19
If the calibration conditions (voltage and current) match the reference meter readings, click the Yes button.
11. The gain register values automatically update when the calibration is complete.
12. To confirm results click the Quick Measurements tab to bring it forward, and click the START button.
13. Results should match the reference meter readings.
Figure 20. Calibration Tab with Gain Register Updated After Gain Calibration
Figure 21. Quick Measurements Tab Showing Results that Should Match the Reference Meter