Note: Pay attention to the fluid

level information on the display:

Low level

High level

(see "Fluid level" in section " Working with Wireless

Controller (WiCo)").

Drain valve

Drain port

- To drain the fluid from the bath, connect a hose to the drain

port and turn the drain valve in counter clockwise direction
with a straight screwdriver.

• Fluids

IKA

®

Designation

Operating temperature range

(°C)

Safety temperature

(°C)

Flash point

(°C)

EG19_N10.100 (H

2

0 + Glycole)

- 10 ... 100

100

125

EG34_N20.100 (H

2

0 + Glycole)

- 20 ... 100

100

125

EG40_N25.100 (H

2

0 + Glycole)

- 25 ... 100

100

125

EG45_N30.100 (H

2

0 + Glycole)

- 30 ... 100

100

125

SI_P20.275 (Silcone Oil)

20 ... 275

275

300

SI_N30.145 (Silicone Oil)

- 30 ... 145

145

170

SY_N10.100 (Synthetic Oil)

- 10 ... 100

100

125

H

2

O (Water*)

5 ... 95

95

Customized 1
Customized 2

WARNING

Fig. 10

Note: Please note the permissible temperature range of

hoses. For hot fluids we recommend the IKA

®

LT 5.20

hoses. When the external system is not necessary, please
seal the pump connectors IN and OUT with the existing

union nuts and stoppers.

• Filling and draining

- Before filling the fluid into the bath, open the front

cover as indicated in following figure.

- Check and make sure that the drain valve is closed (the

stop position in clockwise direction, see Fig. 11).

Note: Please note information in chapter “Commis-

sioning”.
- Connect the mains plug and turn on the device with

mains switch (2).

- The low level icon ( ) appears on the display of the

Wireless Controller (WiCo).

Hose olive

Union nut

Fig. 9

- Open the lid (16) and fill fluid to the bath.

Stop position

Fig. 12