DISPLAY

In the “Display“ menu the user can specify what infor-

mation will be displayed on the screen.

GRAPH

In this menu, you can set the options for the time-tem-
perature diagram.

Automatic:

The scaling of the temperature axis (Y-axis) is automati-
cally determined, depending on the target temperature
and the actual internal and external temperature.
The time axis (X-axis) is permanently scaled to 30 min-
utes.

Manual:
Axis Assignment:

The temperature values to be displayed can be selected.

Axis scaling:

Scaling of the time (X) and temperature axis (Y) can be
selected or set.

PROGRAMS

Under programs, 10 user-defined temperature-time pro-
files can be created. In addition, the pump speed and
behaviour of the external switching outputs (MODULE

M1 and M2) can be defined. A program can consist of

up to 10 segments.
Once a program has been selected, the following op-
tions are available:

Start: Starts the program upon request of loop mode.
Infinite loop: Upon completion of the last segment,

the program continues with the first segment until the
user ends the program by stopping a device function.

Loop number: Indicates the total number of loop cy-

cles until program end.

Note: At the end of the program all device functions

are switched off.

Edit: Edit/change program parameters.
Seg No.: Segment number.
Ctrl. Sensor (int/ext): Determines whether control is

through the internal (int) or external (ext) temperature
sensor.

Temp.: Target temperature.
Ctrl.Mode (Time / +/- x.x K): In "Ctrl.Mode Time"

the target values and settings of the segment are
valid for the duration indicated in the column "Time

hh:mm".

Afterwards, the next program segment is automatically
executed.

In "Ctrl.Mode +/- x.xx K", the hysteresis (tolerance)

of the actual temperature to the target temperature is
set (e.g. +/- 0.1 K). The target values and settings of the
segment are valid until the actual temperature reaches
the target temperature +/- hysteresis for the first time.
Afterwards, the next program segment is automatically
executed.

Pump rpm: Target speed of pump.
M1 (ON / OFF): MODULES M1-Valve:

OFF: M1 valve in initial state

ON: M1 valve in inverted initial state.
Note: The initial state of the M1 valve is defined in

"MODULES M1-Valve" in "Default Status" as

"Open" or "Closed".
M2 (ON/OFF): MODULES M2-Output switch

OFF: M2-output switch in initial state.

ON: M2-output switch in inverted initial state.
Note: The initial state of the M2-Output switch is de-

fined in "MODULES M1-valve" in "Default Status"

as "Open" or "Closed".
Delete: Deletes the selection highlighted in yellow (seg-

ment or program).

Insert: Inserts a new program segment after the se-

lected segment.

Save: Saves changes.
Display: Temperature-time overview displays for the

program with segments of the selected program.

Note: If hysteresis is set as "Ctrl.Mode +/-x.xx K" for

one or more segments in the program, the duration of
the program cannot be determined.
Press and turn the knob (J) to display the segment de-

tails.

Cancel: Cancels the process.
Note: Once the program has been started, the program

no., segment no. (active/total) and the remaining dura-
tion of the segment or hysteresis are displayed in the
graph.

MODULES:

In "Modules" the multifunction connection (19) out-

puts and inputs can be configured.
The outputs M1 and M2 can be controlled via "PRO-

GRAMS".

M1-Valve:

An output to control an external valve, which can be
built into the external pump circuit or the cooling coil
inlet, for example.

ON:

Switches the external valve to the active state (inverted
initial state).