• Cooling coil

When the set temperature close to the ambient temperature, the cooling coil should be used for counter-cooling due
to the self-heating.
Connect the water supply and return line hoses to the cooling coil connectors (see Fig. 2).

For bath temperatures > 95 °C do not use the cooling coil with water.
For bath temperatures > 60 °C make sure that the flow rate through the cooling coil is high enough, to avoid exces-
sive calcification.

• Charging the Battery Pack RB 1 (rechargeable battery)

The battery pack of the WiCo can be charged by any of the following means:

- On the station

- Via a USB cable at the PC or station

- Via an OS 1.0 power supply unit.

1

• Changing battery to WiCo

Please comply with the relevant safety instructions in the "Safety instructions" section for
the RB 1 battery pack!

DANGER

Fig. 13

• WiCo – holder WH 10

WiCo – holder WH 10

WiCo holder WH 10

Fig. 14

WiCo holder WH 10

H 44 Boss head clamp

H 38 Holding rod

* Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.

High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0,1 g soda (Na

2

CO

3

,

sodium carbonate) per liter of water.