IKA HBC 10 control User Manual

Page 44

Advertising
background image

44

Note: When the "PROGRAMS" is started, the M1 seg-

ment settings have higher priority.

Default Status:

Defines the default state (OFF) of the external valve as
"Open" or "Closed".
Refill:

Uses the external M1 valve to automatically check the
level when operating with water.
Observe "Default Status" setting.
Low level: Sets the valve switch-on point (ON).
High level: Sets the valve switch-off point (OFF).
Note: Prior to using the "Refill" option check the func-

tion of the buoyage (4).

The "Refill" option has higher priority than the program

M1 segment settings.

M2-Output:
Alarm:

Activates the alarm output (switch contact).

Error: In the event of an error the "M2-Output —

Switch" is activated (ON, inverted initial state).
Warning: In the event of a warning the "M2-Output

— Switch" is activated (ON, inverted initial state)

When "Warning" is activated, "Error" is simultane-

ously activated.

Default Status: Defines the default state (OFF) of the

"M2-Output Alarm" as "Open" or "Closed".
Switch:

Activates the switch output.

ON: Switches output into the active state (inverted ini-

tial state).

Note: When the "Program" is started, the M1 seg-

ment settings have higher priority.

Default Status: Defines the initial state (OFF) of the

switch output as "Open" or "Closed".

M3-Input:
Ext. Standby:

External standby input to stop the device functions Tem-
per and Pump.

Activated: Activates the ext. standby function. Device

functions are stopped in the event of an inverted initial
state (ON) at the input.

Default Status: Defines the default state (OFF) of the

input as "Open" (high level) or "Closed" (low level).

OUTGAS

In the option, the user is allowed to set the outgas speed
in range of 1000 rpm to 4600 rpm and outgas interval
from 10 seconds to 240 seconds.
This function can be used when filling external devices
such as laboratory reactors.

SAFETY

Time out:

In the menu "Time out", the user can determine a

time limit in the event of a communication breakdown
between the station and the Wireless Controller. The

station continues to work with the preset target values
until the preset time value has expired. Following that,
the station runs using the preset safety temperature and
safety speed.

Note: The initial time out is 30 seconds and the user

can define up to 60 minutes for this time limit.

Speed:

In the “Speed“ menu, the user can specify a appropri-

ate and safe speed for specified temperature.

Note: The factory setting of the safe speed is 1000 rpm

and is activated after the time limit is set (see "Time

out").
Temperature:

In the “Temperature“ menu, the user can specify a

temperature that is appropriate and safe for the circu-
lating task.

Note: The initial safe temperature is 30 ºC and is acti-

vated after the time limit is set (see "Time out").

Password:

In the “Password“ menu, the user can protect the

Wireless Controller settings using a 3 digits password.
When accessing the menu screen, the user is prompted
for a password. 000 deactivates the password prompt
(factory setting).

SETTING

Languages:

The “Language“ option allows the user to select the

desired language.

Display:

The ”Display“ option allows the user to change the

background color and brightness of the working screen.

Sound:

The ”Sound” option allows the user to activate/deacti-

vate the key tone.