IKA RET control-visc User Manual

Page 25

Advertising
background image

25

Operation with the PT1000 double

sensor is indicated on the digital

display by the

symbol. One

temperature sensor is located in the

heat carrier fluid and the other is

located in the reaction medium. The

set target temperature (2) refers to

the temperature of the medium. The

higher temperature is automatically

attributed to the heat carrier fluid and

shown on the display as an additional

current temperature with the symbol

(3).

The actual temperature of the reaction

medium is shown as the current

temperature (5). The heating plate

temperature (and therefore the heat

carrier fluid temperature) are limited by the set heating plate

safety temperature limit (7). The reaction medium temperature

is limited by the set temperature limits (carrier fluid/reaction

temperature limit).

see section entitled “Limits”

Limits

In order to ensure that neither the reaction fluid nor the carrier

fluid overheat, the user can set separate temperature limits.

• Scroll to the “Heating” icon

in the menu

• Open the submenu by

pressing the rotary knob (D)

• Open the “Limits” submenu

followed by “Medium
limit” and then set the
temperature limit for the
reaction fluid.

• Confirm the set limit by

pressing the rotary knob (D)

• Open the “Carrier limit”

submenu and set the limit
temperature for the carrier
fluid.

• Confirm the set limit by

pressing the rotary knob (D)

Regulation of the temperature of the medium
using PT1000 double temperature sensor

• Open the “Limit

information” submenu.
All set limit temperatures,
including the safety limit
temperature of the heating
plate, as displayed on one
screen.

• Close the submenu by

pressing the rotary knob (D)

Calibrating the temperature sensor

In order to use tolerances to limit temperature deviations, the

user can calibrate the temperature sensor in conjunction with the

device.

2-point calibration:

Calibration using two temperatures

3-point calibration:

Most accurate way of calibrating the temperature sensor at three

different temperatures

Note:

A calibration carried out by the user is indicated on the display

with the symbol (11) shown in red.The calibration of the PT100

and PT1000.50 double

sensor are carried out

separately.

• Scroll to the “Heating” icon

in the menu

• Open the submenu by

pressing the rotary knob (D)

• Open the “Temperature

sensor” submenu and
then “Calibration”, finally
confirming by pressing the
rotary knob (D)

• Select either 2-point

calibration or 3-point
calibration and confirm by
pressing the rotary knob (D)

In this function, the speed can be pre-set before starting the

calibration.