IKA RET control-visc User Manual

Page 27

Advertising
background image

27

The current control variable is saved as the reference 100% ΔP
and shown on the digital display. The change in the viscosity is
then shown in %. Depending on whether the viscosity increases
or decreases, the percentage rises or falls above or below 100%
respectively.
• Scroll to the “Stirring” icon

in the menu

• Open the submenu by

pressing the rotary knob
(D)

• Open the “Torque trend

measurement” submenu
and confirm by pressing
the rotary knob (D)

• Acknowledge the informa-

tion screen displayed

Weighing module

The weighing module is a function which allows the user to

perform simple weighing

tasks.

• To start the weighing pro-

cess, press the (G) button

on the front membrane or

alternatively select Weig-

hing from the menu. Press
the rotary knob (D) to start
the weighing function.

Note:

The heating and stirring

functions must be deacti-

vated.

• It may take several

moments for the system to

stabilise.

Once the system is stable,

a “g” will be shown

alongside the measured

value.

Weighing - Calibration

• Open the “Calibration”

submenu and confirm by
pressing the rotary knob
(D)

• Enter the calibration

weight and confirm by
pressing the rotary knob
(D)

• Place the calibration

weight on the mounting
plate and wait until the
device indicates that the
calibration process has
been completed.


This process needs to be
repeated two or three
times before closing cali-
bration mode

Once the calibration
process has been
completed successfully, the
weighing module is ready
for use.

Regularly re-calibrate the
device.

Timer / Counter

The Timer function allows the user to stop the heating process

after a certain time has lapsed. The Timer function can be set for
up to 99 hours, 59 minutes and 59 seconds.

Note:

The time starts to count down when starting both functions

(heating and/or stirring).


Once the heating function is activated, the timer function only
refers to heating. The stirring function continues to operate as
previously, thereby avoiding a delay in boiling in the liquid. The
motor is only switched off when the heating function is not
active. The stirring function is stopped if the heating function is
not started during the timer cycle and the time runs out.

NOTE:
By entering a value of 00:00:00 (hh:mm:ss), the counter is

activated. The counter shows the time elapsed during the current
experiment.