IKA RET control-visc User Manual

Page 28

Advertising
background image

28

Setting of the timer

• To start the timer function,

press the (H) button on

the front membrane or

alternatively select “Timer”

from the submenu.

The timer settings are

opened by clicking on
“Set”. The timer can be set
by turning the rotary knob
(D).

• Set the hours and press the

rotary knob (D)

• Set the minutes and press

the rotary knob (D)

• Set the seconds and press

the rotary knob (D)

The selected timer (count-
down) time is shown on
the display.


Safety settings

The user has several options

in the safety settings menu,

and the device functions can

be set according to the user’s

individual requirements.

• To start, select “Safety” in

the submenu and confirm
by pressing the rotary knob
(D).

The “Timer” function is used to
define how long an interruption
in communications between the
magnetic stirrer and the wireless
controller box can be before there
is a reaction in the magnetic stirrer.

If the set time limit is exceeded,
the pre-selected temperature and
speed values become active.

Please refer to the separate in-
struction manual for the wireless
controller box for more details.

In the other menu items, the
user can enter limit values for the
temperature and speed.

• Select “Password” from the

submenu and confirm by
pressing the rotary knob
(D).
Enter a combination of
digits as your password and
confirm by pressing the
rotary knob (D).

This information screen
informs the user that they
must confirm the hot
plate safe temperature
value when starting up the
device.

• Safety circuit confirmation

at device start-up: the hot
plate safe temperature
value must be confirmed by pressing the rotary knob (D).

pH measurement

The device has an integrated pH measuring feature for

determining the pH of different liquids.

The pH sensor must be calibrated before being used to attempt a

pH measurement.

The calibration is used to adjust the pH probe and the device

so that they work together correctly. As part of the process,

the neutral and pH gradient are specified for the measurement

chain. To complete the calibration, use pH4, pH5 and pH7 buffer

solutions in accordance with DIN 19266.

Note:

a pH calibration can only be carried out with an inserted

temperature sensor.

• Scroll down to the “pH

probe” icon in the menu

• Open the submenu by

pressing the rotary knob (D)

• Open the “Calibration”

submenu and confirm by
pressing the rotary knob (D)

• Place the pH probe and

temperature sensor in the
required buffer solution.

• Once the value is stable,

confirm it by pressing the
rotary knob (D).

• Follow the instructions on

the display and place the
pH probe and temperature
sensor in the next buffer
solution.