Pin connector serial rs 232 interface and usb – IKA RET control-visc User Manual

Page 30

Advertising
background image

30

Power supply

connection

PT 100 / PT 1000 plug

RS 232

BNC connector (only for pH probe)

Pt1000-2

5

3

1

4

2

Pt1000-1

Pt100

5

3

1

4

2

9-pin connector

Serial RS 232 interface and USB

Configuration

The RS 232 serial interface can be used to operate the device
using a computer and a suitable user program (labworldsoft).
In order to increase safety when operating the RET using a

PC, a watchdog function can be activated which monitors

the continuous data flow (see section entitled: “Watchdog

function”).

• The functions of the interface circuit between the laboratory

device and the automation system are a selection from the
signals specified in the EIA standard RS232 as per DIN 66020
Part 1. The assignment of the different signals can be seen in
the image.

• Standard RS 232, corresponding to DIN 66259 Part 1 is valid

for the electronic characteristics of the interface circuits and
assignment of signal states.

• Transmission process: asynchronous character transmission in

start-stop operation.

• Transmission type: full duplex
• Character format: character composition according to data

format in DIN 66022 for start-stop operation. 1 start bit,
7 character bits, 1 parity bit (even), 1 stop bit.

• Transmission speed: 9,600 bit/s
• Data flow control: hardware handshake RTS/CTS

RTS: (PIN 7) LOW (positive voltage)

/ COMPUTER can transmit

RTS: (PIN 7) HIGH (negative voltage)

/ COMPUTER cannot transmit

CTS: (PIN 8) LOW (positive voltage)

/ COMPUTER ready-to-receive

CTS: (PIN 8) HIGH (negative voltage)

/ COMPUTER not ready-to-receive

• Access method: data transmission from the device to the

computer only occurs after a request from the computer.

Command syntax

The following points should be noted for the instruction set:
• Commands are generally sent from the computer (master) to

the lab device (slave).

• The lab device only responds to requests from the computer.
• Commands and parameters, as well as consecutive parameters,

must be separated by at least o n e space (code: hex 0x20).

• Each individual command including parameters and data and

all responses are completed with CR LF (code: hex 0x0D and
0x0A) and can have a maximum length of 10 characters.

• The dot is used for decimal separators in a floating-point value

(code: hex 0x2E).

The details given above generally comply with the recommendations
of NAMUR (NAMUR recommendations for the design of electrical
plug-in connectors for analogue and digital signals in laboratory
MSR devices. Rev. 1.1).

Summary of available NAMUR commands

Abbreviations used:
X,y = numbering parameter (integer)
m = variable value, integer
n = variable value, floating-point number
X = 1 medium temperature(external temperature sensor)
X = 2 hot plate temperature
X = 3 hot plate safety temperature
X = 4 speed
X = 5 viscosity trend
X = 7 heat transfer medium temperature
X = 80 pH value
X = 90 Weight value
X = 54 Error 5 response time in seconds (180 <= n <= 1200)
X = 55 Intermittent mode cycle time in seconds

( 10 <= n <= 600 )

X = 56 Intermittent mode, duration of interruption in seconds

( 5 <= n <= 60 )