NAMUR commands

Function

Display

(additional)

IN_NAME

Title request

IN_PV_X X=1;2;3;4;5;7;80;90 Current value reading

IN_SOFTWARE

Software ID number, date, version request

IN_SP_X

X=1;2;3;4;7;8;12;
42;54;55;56;

Set target value reading

IN_TYPE

Lab device identification request

OUT_NAME name

Output of identification name. (Max. 6 characters; default: IKAMAG)

OUT_SP_12@n

Setting WD safety limit temperature with set value echo

OUT_SP_42@n

Setting WD safety limit speed with set value echo

OUT_SP_X n

X=1;2;4;7;
54;55;56

Setting of target value to n

OUT_WD1@m

Watchdog mode 1: if event WD1 should occur, the heating and stirring functions are
switched off and ER 2 is displayed. Set watchdog time to m (20 - 1,500) seconds,
with watchdog time echo. This command launches the watchdog function and must
be transmitted within the set watchdog time.

OUT_WD2@m

Watchdog mode 2: if event WD2 should occur, the speed target value is changed to
the WD safety speed limit and the temperature target value is changed to the WD
safety temperature limit value. The warning WD is displayed. The WD2 event can be
reset with OUT_WD2@0 - this also stops the watchdog function. Set watchdog time
to m (20 - 1,500) seconds, with watchdog time echo. This command launches the
watchdog function and must be transmitted within the set watchdog time.

RESET

Switches off the device function.

START_X X=1;2;4;5;7
X=80; 90

Switching on of device - (remote) function
Activation of the previous watchdog function. If no target value is entered within
60 seconds, the (old function) heating is switched off.

Remote

STOP_X X=1;2;4;5;7
X=80; 90

Switching off of device - (remote) function The variables set with OUT_SP_X remain
saved. Includes command RMP_STOP.

Remote