“Watchdog” functions; monitoring of the serial data flow

If, once this function has been activated (see NAMUR commands),
there is no retransmission of the command from the computer
within the set time (“watchdog time”), the heating and stirring
functions are switched off in accordance with the set “watchdog”
function or are changed to the set target values.
The data transmission may be interrupted by, for example, a crash
in the operating system, a power failure in the PC or an issue with
the connection table between the computer and REC control.

“Watchdog” - mode 1

If there is an interruption in data communications (longer than
the set watchdog time), the heating and stirring functions are
switched off and ER 2 is displayed.

“Watchdog” - mode 2

If there is an interruption in data communications (longer than
the set watchdog time), the speed target value is changed to
the WD safety speed limit and the temperature target value is
changed to the WD safety temperature limit value. The warning
WD is displayed.

Connection possibilities between the RET control

and external devices

The following adapters and adapter cables are available from IKA

®

for connecting the RET control to external devices.

PC 1.1 cable

The PC 1.1 cable is used to connect the 9-pin plug to a computer.

Accessories

Troubleshooting


The device will indicate that there is a fault by displaying an error
message “Er X” (see table) on the screen.
Heating will be switched off permanently (exception: WD - see
table).

The stirring function continues to operate at the pre-set speed if
there is a fault in the heating circuit (exception: WD - see table).
In order to reset the device after a fault, it may need to be discon-
nected from the power supply for several seconds. The period the
device is switched off may also need to be longer if the tempera-
tures need to be counterbalanced (residual heat display may no
longer be active; see “Residual heat display”).

The safety circuit test takes place within 20 seconds of switching
on the device; after the test, the other monitoring functions are
activated.

If a fault cannot be eliminated directly, you must perform a

RESET (see “Settings” menu). If the faults still cannot be elimi-

nated, the appliance must be inspected by a technical service

technician.


• Magnetic rods:

ø 6 mm; max. length 15 mm

ø 7 mm; max. length 60 mm

ø 10 mm; max. length 80 mm

• RS 1

Set of magnetic stirring bars

• RSE

Stirring bar remover

• H 1000

Beaker

1 l

• H 1500

Beaker

1.5 l

• H 3000

Beaker

3 l

• H 5000

Beaker

5 l

• H 8000

Beaker

8 l

• H 135.3

Flask carrier

• H 135.310

Quarter 20 ml reaction vessel

• H 135.311

Quarter 30 ml reaction vessel

• H 135.312

Quarter 40 ml reaction vessel

• H 135.313

Quarter 4 ml reaction vessel

• H 135.314

Quarter 8 ml reaction vessel

• H 135.315

Quarter 16 ml reaction vessel

• H 135.4

Reaction block for 100 ml round flask

• H 135.410

Insert for 10 ml round flask

• H 135.411

Insert for 25 ml round flask

• H 135.412

Insert for 50 ml round flask

• H 135.5

Reaction block for 500 ml round flask

• H 135.510

Insert for 200 ml round flask

• H 135.511

Insert for 250 ml round flask

• H 135.512

Insert for 300 ml round flask

• H 135.6

Reaction block for 1,000 ml round flask

• H 135.610

Insert for 500 ml round flask

• H 16 V

Support rod

• H 16.1

Extension

• H 38

Holding rod

• H 44

Boss head clamp

• PT100.50

Temperature sensor

• PT100.51

Temperature sensor, glass coated

• PT1000.50

Double temperature sensor

• PT1000.51

Double temperature sensor, glass coated