61

8.4.2

Self test

The self test function of the Mastercycler pro allows you to perform a quick check regarding the following
attributes of the thermoblock without requiring an external measurement system:

• Heating and cooling rates
• Proper functioning of the temperature control circuit
• Temperature homogeneity in each of the three zones of the block and across the entire block

After a successful self test, a certificate can be generated in PDF format.

A separately available USB key is required for performing the self test (see Accessories on p. 85). Please
refer to the documentation of the self test USB key for a detailed description of the self test function.

8.5

Managing and backing up data

8.5.1

Backing up data

The function keys in the above view show the options available to the administrator for transferring data.
Standard users only have the option of performing backups here.

 Insert an MMC memory card or a USB storage medium into the right-hand side of the control panel.