Full-Scale Range Adjust (RANGE):
Allows the user to adjust the reading at full-scale pressure. The reading can be
adjusted by +/-10% full-scale.

NOTE: The pressure reading must be within 5% of the full-scale value in order
to make Range adjustments. For example, a 1000 psi gauge would need

to be reading between 950 psi and 1050 psi in order to adjust the Range

feature. If the reading is not within 5% of full scale, NOAdJU (No

Adjustment) will be displayed when trying to adjust the Range.

Adjusting the Full-Scale Range:

Step 1: Ensure that the pressure reading is within 5% of full-scale.

Step 2: Press to enter the main setup menu.

Step 3: Press or to cycle through the options until AdVSET is

displayed.

Step 4: Press to access the Advanced Settings menu.

Step 5: Press or to cycle through the options until RANGE is displayed.

Step 6: Press to enter the Range adjust mode.

Step 7: Press to increase the reading, or press to decrease the reading.

Step 8: Press to accept the adjusted reading and return to the advanced

setup

menu.

Step 9: To exit the Advanced Setup menu, press or until EXIT is

displayed and press to exit to the main setup menu.