Fast Calibration (CALIBR):
Allows the user to perform a fast calibration of the zero and span.

NOTE: Both zero pressure and full-scale pressure must be applied to the gauge
in order to complete the quick calibration process.

Step 1: Press to enter the main setup menu.

Step 2: Press or to cycle through the menu options until AdVSET is

displayed.

Step 3: Press to enter the Advanced Settings menu.

Step 4: Press or to cycle through the options until CALIBR is

displayed.

Step 5: Press to access the CALIBR mode. The gauge will display the

word ZERO, prompting the user to perform the zero pressure quick

calibration.

Step 6: Ensure no pressure is applied to the gauge. Press to zero the gauge.

The gauge will then display the word SPAN, prompting the user to

perform the full-scale pressure quick calibration.

Step 7: Apply full pressure to the gauge and press to complete the quick

calibration and return to advanced setup menu.

Step 9: To exit the advanced setup menu, press or until EXIT is

displayed and press to exit to the main setup menu.