CiDRA SONARtrac PW VF-100 User Manual
Page 224
20838-01 Rev 09
Page A-2
Division 2
Each Transmitter includes two 5mm x 20mm cartridge fuses on the
Power Entry board next to the terminal block connections for mains
power. Different models require different fuses, though all are rated
for 250VAC. In all cases, the fuse information is included on a label
inside the Transmitter enclosure. Below are the fuse requirements by
Transmitter model number.
Transmitter Model #
Rating
Part
Number
Vendor p/n Notes
TB8-xx-05-x1-xx
1 A
52105-10
Bussmann
S506-1A
1
TB8-xx-06-x1-xx
3.15 A
52105-15
Bussmann
S506-
3.15A
2
TB8-xx-05-x2-xx
0.5 A
E50382-05 Littelfuse
215.500
3
TB8-xx-06-x2-xx
1.6 A
E50382-10 Littelfuse
21501.6
4
Where “x” = any alphanumeric character
Notes:
1) Or any UL and VDE (IEC60127-2-3) recognized 5x20 time-delay
fuse rated 1A, 250V, with a minimum 35A breaking capability.
2) Or any UL and VDE (IEC60127-2-3) recognized 5x20 time-delay
fuse rated 3.15A, 250V, with a minimum 35A breaking capability.
3) Or Littelfuse 215.500P; or Bel Fuse 5HT500 or 5HT500-R. These
are 1500A breaking ceramic fuses and are the only fuses
permitted under the ATEX Zone 2 certification. For Ordinary
Location Transmitters (TB8-xx-05-x2-01) ONLY, any UL and VDE
(IEC60127-2-3) recognized 5x20 time-delay fuse rated 0.5A, 250V,
with a minimum 35A breaking capability.
4) Or Littelfuse 21501.6P; or Bel Fuse 5HT1.6 or 5HT1.6-R; or
Schurter 0001.2506; or Ferraz Shawmut UDA1.60; or
Cooper/Bussmann S505-1.6A or S505-1.6-R. These are 1500A
breaking ceramic fuses and are the only fuses permitted under the
ATEX Zone 2 certification. For Ordinary Location Transmitters
(TB8-xx-06-x2-01) ONLY, any UL and VDE (IEC60127-2-3)