2 armored cable installation -202, Figure 14, Nema 4x rated connector -20 – CiDRA SONARtrac PW VF-100 User Manual

Page 86: Table 3

Advertising
background image

20832-01 Rev 05

Page 6-20

The following table lists the cable connector pin-out for cables
equipped with the NEMA 4X rated connector (Figure 14).

Table 3

Sensor to Transmitter Cable Terminal Connections NEMA 4X

Connector

Figure 14

NEMA 4X Rated Connector

Wire Pair # Transmitter

Terminal #

Function

SENSOR HEAD
CONNECTOR PIN
NUMBER

1 Wht/Blk

1

Sensor #1 Input

1 - Wht

13 - Blk

2 Wht/Blk

2

Sensor #2 Input

2 - Wht

14 - Blk

3 Wht/Blk

3

Sensor #3 Input

3 - Wht

15 - Blk

4 Wht/Blk

4

Sensor #4 Input

4 - Wht

16 - Blk

5 Wht/Blk

5

Sensor #5 Input

5 - Wht

17 - Blk

6 Wht/Blk

6

Sensor #6 Input

6 - Wht

18 - Blk

7 Wht/Blk

7

Sensor #7 Input

7 - Wht

19 - Blk

8 Wht/Blk

8

Sensor #8 Input

8 - Wht

20 - Blk

9 Wht/Blk

9

Spare

– unused

---

---

10 Wht/Blk

10

Wht

– RS 485 Hi / Blk – RS485 Low

12 - Wht

24 - Blk

11 Wht/Blk

11

Wht

– ‘-12 V’ / Blk – Gnd

9 - Wht

21 - Blk

12 Wht/Blk

12

Wht

– ‘+12 V’ / Blk – Gnd

10 - Wht

22 - Blk