20838-01 Rev 09

Page J-8

Item
No

PROBLEM

POSSIBLE
CAUSE

DIAGNOSTIC
METHOD

DIAGNOSTIC
RESULT

COMMENTS

13

The signal
output from
the meter
has spikes
during no
flow
conditions or
drops out
during flow
conditions

Need to use Spike
Filter

See Filter section of
Manual in order to turn
on filtering

Yes

– With filtering

Enabled, signal is less
noisy

Use this as new transmitter configuration

No

– With filtering

enabled, signal is still
noisy

Acquire Snapshot, Upload to secure Website,
and contact Customer Support.

14

Flow does
not seem
correct

Improper pipe size
entered

Verify pipe size input in
transmitter ‘PIPE SIZE’
in BASIC CONFIG
menu

Yes

– correct size

entered

See below

No

– incorrect pipe size

entered

Re-enter correct pipe size

Improper calibration
coefficients entered

Verify coefficients in
transmitter
‘CALIBRATION’ in
BASIC CONFIG menu
‘match sensor band
coefficients (on label in
transmitter or on sensor
band cable)

Yes

– correct values

entered

See below

No

– incorrect values

entered

Re-enter correct coefficients

All sensors are not
turned on

Verify sensors are
turned on in ‘SENSOR
SETUP / ENABLE’ in
CUSTOMIZE menu

Yes

– all sensors are

on

See below

No

– some or all are

turned ‘Off’

Turn on all sensors in Customize menu
unless some were purposely turned off.
If problem persists, Acquire Snapshot, Upload
to secure Website, and contact Customer
Support.