Flow meter faults (page 5) – CiDRA SONARtrac PW VF-100 User Manual
Page 271
Advertising

20838-01 Rev 09
Page J-12
Flow Meter Faults (page 5)
Item
No
PROBLEM
POSSIBLE
CAUSE
DIAGNOSTIC
METHOD
DIAGNOSTIC RESULT
COMMENTS
18
The Flow
Reading is
erratic
The meter is
mounted too close to
an elbow or
obstruction
If possible move meter
to a longer straight run
of process pipe.
Yes
– The reading is less erratic
No
– The reading is still erratic.
Acquire Snapshot, Upload
to secure Website, and
contact Customer Support.