2 transmitters with pluggable terminal blocks, 2 transmitters with pluggable terminal blocks -15, Figure 9 – CiDRA SONARtrac PW VF-100 User Manual

Page 81: Transmitter output terminals -15, Table 2, List of transmitter signal outputs -15

Advertising
background image

20832-01 Rev 05

Page 6-15

6.3.4.3.2

Transmitters With Pluggable Terminal Blocks

The following figure provides a close-up of the transmitter output
terminals (Section #1 of the terminal board) with their functions listed
in the following table. These outputs can be connected as appropriate
to permit communications between the transmitter and other
equipment.

The recommended torque for the terminal screws is 4.4 to 5.3 lb

f

-inch

(0.5 to 0.6 Nm).

The recommended torque for the screws that secure the terminal
blocks to the base is 3.5 to 4.4 lb

f

-inch (0.4 to 0.5 Nm).

Figure 9

Transmitter Output Terminals

Terminal Label

Type

Comment

CUR 1

4-20mA #1 Output

Internal (self) or external (loop) powered,
HART communication capable

CUR 2

4-20mA #2 Output

Internal (self) or external (loop) powered

PULSE

Pulse output

Solid-state relay closure

ALARM

Alarm output

– High /

Low Output

Solid-state relay closure

COMM

RS-485 or RS 232
Digital Output

MODBUS communications on those units
with this option

Foundation

Digital

Optional

Foundation Fieldbus™ terminals

SHD

---

Shield Connections

Table 2

List of Transmitter Signal Outputs

Optional Foundation
Fieldbus™ terminals

Advertising