Description

RS485 i
Connection for RS-485

A: Active bus terminator
B: Without bus terminator
C: Passive bus terminator

Galvanic isolation (max. 60 V DC, 25 V VAC against GND)

Pin assignment:

The bus terminator integrated in the I/O module (120 Ω) between D+ and D can be
disabled via DIP switches S2-4. The BUS address is 1. BUS operation with more
than two subscribers is not possible.

STATUS-LED c

Fig. 5

RS-485 bus terminator

1

2

3

4

5

562 Ω

120 Ω

562 Ω

DIP S2-4

RS485

D+

D-

+5 V

0 V

I/O-Modul

A

1

2

3

4

5

562 Ω

120 Ω

562 Ω

DIP S2-4

RS485

D+

D-

+5 V

0 V

I/O-Modul

B

1

2

3

4

5

562 Ω

120 Ω

562 Ω

DIP S2-4

RS485

D+

D-

+5 V

0 V

I/O-Modul

C

Pin Name

1

Connect pull-up resistor (562 Ω against +5 V) with D+ if necessary

2

D+

3

D-

4

Connect pull-down (562 Ω against GND) with D- if necessary

5

COM

Color

Status

Meaning

Red

illuminates

Unit not functional or defective

Red

flashes

Not ready for operation, communication to leak
detector is not available

Cyan

illuminates

Ready for operation; communication to leak
detector available

Green

Flashes quickly

Boot loader active, ready for software update

Green

Flashes slowly

Data reception on RS232

Yellow

Flashes slowly

Data reception on RS485

off

No operating voltage