4.1
Construction of the I/O module
PLC IN a
Digital inputs
Galvanic isolation (max. 60 V DC, 25 V VAC against GND)
Max. permissible input voltage: U = 35 V
Active signal: U = 13 ... 35 V (typically 24 V), I = approx. 7 mA
Inactive signal: U < 7 V (typically 0 V), I = 0 mA
Pin assignment: The input pins PLC-IN 1 to PLC-IN 10 can be configured freely, see
the following example.
Fig. 1
Front view
a
PLC IN, digital inputs
b
RS232
c
Status LED
d
ANALOG OUT, analog outputs
e
PLC OUT, digital outputs
f
DIP switch cover and fuse
g
LD, connection of leak detector
h
ANALOG IN, analog inputs
i
RS485
j
24 V OUT, 24 volt output
a
b
c
d
e
f
g
h
j
i
24 V
OUT
RS485
ANALOG
IN
PLC IN
LD
RS232
STATUS
S1
FUSE
ANALOG
OUT
S2
PLC OUT