INFICON IO1000 I/O module User Manual

Page 9

Advertising
background image

Description

9

PLC OUT e
Digital outputs

Galvanic isolation (max. 60 V DC, 25 V VAC against GND)
Max. permissible load per output: U = 30 V, I = 0.75 A

Fuses for digital outputs 1 ... 4 and 5 ... 8: 2 x 0.75 A

Pin assignment: The output pins PLC_OUT_1 to PLC_OUT_8 can be configured
freely, see the following example.

Fig. 2

Sample wiring, digital inputs with PLC. Left: Driver outputs. Right: Potential-free contacts

PLC-IN

PLC-IN

1 Dyn CAL / Normal Cal

1 Dyn CAL / Normal Cal

2 SNIFF / VAC

2 SNIFF / VAC

3 STOP

4 ZERO

5 EXT.CAL

6 INT.CAL

7 CLEAR

8

9

10

11 PLC-IN-GND

3 STOP

4 ZERO

5 EXT.CAL

6 INT.CAL

7 CLEAR

8

9

10

11 PLC-IN-GND

24 V-OUT

+

_

0 V

24 V

0–24V

PLC/
SPS

PLC/
SPS