Description

13

FUSE and DIP switch S1, S2 f
Fuses for digital outputs and DIP switches (under the cover)

Fuses for digital outputs 1 ... 4 and 5 ... 8:
2 x 0.75 A (Schurter: 7010.9800.xx)

1 = ON, 0 = OFF

1 = ON, 0 = OFF, + = switching from OFF to ON during operation,
X = random

Connection LD g
Connection for the data cable to the leak detector

24V OUT j
24V output

Pin assignment:

The I/O module is supplied with voltage by the leak detector and requires no
separate power supply. The 24V output is not used for voltage supply to the
I/O module.

The 24V output of the I/O module can be used as an active signal for the PLC inputs
and outputs.

DIP switch S1

LDS3000 HLD6000

Contact
4 3 2 1

Factory setting (default value of the interface
protocol by the leak detector or control unit)

0 0 0 0

ASCII protocol

0 0 1 0

LD protocol

0 0 1 1

Binary protocol

0 1 0 1

LDS1000 protocol

0 1 1 0

Normal protocol

0 0 0 1

Simple protocol

0 1 0 0

DIP switch S2

Contact
4 3 2 1

Activate boot mode for software update

X + 0 0

Disable bus terminator 120 Ω for RS-485

1 X 0 0

Pin Name

+

+24 V

-

GND