Installation and removal

15

5

Installation and removal

5.1

Mount I/O module on DIN-TS35 top hat rail

1 Hook unit on top hat rail at bottom.

2 Press unit onto top hat rail at top.

5.1.1

Establish connections

Connecting I/O module with leak detector

The I/O module communicates via data cable with the leak detector and is supplied
with voltage by the data cable.

1 Connect I/O module (connection “LD”) via data cable with leak detector

(connection “I/O Anybus”).

2 Connect I/O module via desired interfaces with external controller:

– RS232 (RS-232 interface)

– RS485 (RS-485 interface)

– Analog in (analog inputs)

– Analog out (analog outputs)

– PLC IN, digital inputs

– PLC OUT, digital outputs

Fig. 6

Mounting of I/O module