4 hand probe p60, 1 general information, Hand probe p60 – INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual

Page 19: General information, 4hand probe p60

Advertising
background image

17

ni

n65

e1-a (140

9)

4

Hand Probe P60

4.1

General Information

The Hand Probe P60 is a non-sucking probe. Gas sensing takes place in the
replaceable sensor located in the tip of the probe.

Caution

Do not expose the probe to a hydrogen concentration greater than 0.1% when the
instrument is not powered on, as this could damage or destroy the probe gas
sensor.

Caution

Connection and disconnection of the sensor cable must be done with power off.
Sensor can be damaged if power is on.