10.2

How to Calibrate

INFICON support two types of Calibration Leak. The one with large leaks must be
connected to tracer gas pressure to work. The small leak is connected to a small fillable
gas container. Read the instruction delivered with the leak how to set up and how to
use the leak.

10.2.1

Prepare the Calibration Leak

Leak with container:

1

Fill the gas container to the indicated pressure with tracer gas.

2

Leak is ready to use.

Leak without container:

1

Connect the leak with tracer gas with the pressure according the certificate.

2

Vent the hose from air via the Purge Valve on the leak.

3

Start to use the leak.

10.2.2

Set the Calibration Value

1

Click Settings >> Calibration >> Setup 1.

2

Set the Calibration Unit and Calibration Value (as stated on the calibration
certificate issued for the leak).

3

If using Calibration Leak: Click Settings >> Calibration >> Setup 2 to set
calibration Leak Gas type.

4

Set Leak Gas (as stated on the calibration certificate issued for the leak)