41

ni

n65

e1-a (140

9)

8.5

I·Guide

Summarizing the measuring results. A fixed or moving number of measurements with a
maximum 25 measurement points can be chosen. The instrument must be in
Measuring Mode or Combined Mode for this function to be active. Toggle the Mode
function for Hand Probe.

To use I·Guide

If a fixed number of measurement points is used, then measure according to the
following steps:

1

Press the probe button, or start button on the screen, to begin the first
measurement.

2

Place the probe on the test subject while the time-staple moves. The instrument
registers the results.

3

You might need to wait until the next measurement. The instrument signals Wait.

4

Repeat the procedure for the next measuring point.

When all measurements are made, the sum of all leaks is shown. If the sum of all leaks
is greater or the same as the Reject Level, REJECT is shown. If the sum of all leaks is
under the Reject Level, ACCEPT is shown. And if the sum of all leaks is greater than
Reject Level before all measurements are made then REJECT is shown.

To begin a new measurement action or to stop a current measurement, press the
probe button and hold the button for a moment.

It is possible to measure or search in Combined Mode for a leak without registering the
measurement. A measurement value is only registered when the time-staple is moving.