Notes – INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual
Page 71

69
ni
n65
e1-a (140
9)
Notes
1) OUTNote on RS-232 communication: This system is a DCE device with only RXD,
TXD and GND. Remaining RS-232 signals are not implemented. A DB9-to-DB25
adapter can be used: then connect the DB25 male to this system (DCE) and connect
the DB9 female to the terminal device (DTE).
2) 2A = 24V + total Digital outputs current.
3) Nominal level = [0 ... +24] V
Maximum level = [-50 ... +50]V
Low-high threshold = appr. 8.5V
Input impedance = min 23kohm
Input functions selectable in the menu:
•
Unused
•
Start (I·Guide)
•
Stop (I·Guide)
•
Locating Signal Zeroing
•
External Alert (Error high)
4) Nominal level = [0 ... +24] V
Maximum sourcing current = 500mA
Output impedance = max 4ohm
Output functions selectable in the menu:
•
Unused
•
Reject
•
Accept (only I·Guide)
•
Calibrated
•
Not Ready
•
Gas Signal
•
Not Ready or Gas Signal
•
Error (summary error)
5) (Outputs)
Unused
Reject
Accept
Calibrated
Not Ready
Gas Signal
Not Ready or Gas Signal
Error (summary error)
6) (Inputs)
Unused