Notes

1) OUTNote on RS-232 communication: This system is a DCE device with only RXD,
TXD and GND. Remaining RS-232 signals are not implemented. A DB9-to-DB25
adapter can be used: then connect the DB25 male to this system (DCE) and connect
the DB9 female to the terminal device (DTE).

2) 2A = 24V + total Digital outputs current.

3) Nominal level = [0 ... +24] V

Maximum level = [-50 ... +50]V

Low-high threshold = appr. 8.5V

Input impedance = min 23kohm

Input functions selectable in the menu:

•

Unused

•

Start (I·Guide)

•

Stop (I·Guide)

•

Locating Signal Zeroing

•

External Alert (Error high)

4) Nominal level = [0 ... +24] V

Maximum sourcing current = 500mA

Output impedance = max 4ohm

Output functions selectable in the menu:

•

Unused

•

Reject

•

Accept (only I·Guide)

•

Calibrated

•

Not Ready

•

Gas Signal

•

Not Ready or Gas Signal

•

Error (summary error)

5) (Outputs)

Unused

Reject

Accept

Calibrated

Not Ready

Gas Signal

Not Ready or Gas Signal

Error (summary error)

6) (Inputs)

Unused