INFICON Sensistor Sentrac Hydrogen Leak Detector User Manual

Page 70

Advertising
background image

68

ni

n65

e1-a (140

9)

11

Analog input 0-10V

Nominal level = [+0.05 ... +9.95] V

Maximum level = [-50 ... +50] V

Input impedance = 22kohm

Input LP-filter fc = 2.9kHz

A/D internal resolution: 10-bit

Not used

12

PLC input EXT0

See note 3)

Selectable input function, see
note 6)

13

PLC input EXT2

See note 3)

Selectable input function, see
note 6)

14

Audio line out return (GND)

GND External Speaker

15

Audio line out

AC-coupled

Output is unbalanced

Line_OUT- is connected to GND

Line_OUT+ carries analog audio
symmetrically to GND

Output voltage swing, min = -
1.65V

Output voltage swing, max =
+1.65V

Output impedance = 160ohm @
400Hz

136ohm @ 1kHz

122ohm @ 10kHz

Output LP filter fc = 133kHz

Output External Speaker

16

PLC output EXT0

See note 4)

Selectable output function, see
note 5)

17

PLC output EXT1

See note 4)

Selectable output function, see
note 5)

18

PLC output EXT2

See note 4)

Selectable output function, see
note 5)

19

PLC output EXT3

See note 4)

Selectable output function, see
note 5)

20

RS232: DTE ready

See note 1)

Serial communication

21

GND

GND for I/O

22

RS232: Ring indicator

See note 1)

Serial communication

23

Analog output 0-10V

Nominal range = [0.150 ... +10.0]
V

Resolution: 8-bit

Output current = -65mA (sourcing)
/ +100mA (sinking)

Unlimited capacitive load

No short-circuit protection

Linjear output betvieen to
measuring level. Check value
level, Calcualte RL voltage,
calcualte resulution.

No Locating value.

24

GND

GND for I/O

25

PLC input EXT1

See note 3)

Selectable input function, see
note 6)

Pin

Signal

Specification

Function

Advertising