3 troubleshooting sensors – INFICON Q-pod Thin Film Deposition Monitor User Manual

Page 31

Advertising
background image

4 - 3

IP

N 07

4-

54

7-

P1

B

Q-pod Operating Manual

4.1.3 Troubleshooting Sensors

NOTE: Many sensor head problems may be diagnosed with a DMM (Digital

Multi-Meter). Disconnect the short oscillator cable from the feedthrough
and measure the resistance from the center pin to ground. If the reading is
less than 10 megohms, the source of the leakage should be found and
corrected. Likewise, with the vacuum system open check for center
conductor continuity, a reading of more than 1 ohm from the feedthrough
to the Q-pod indicates a problem. Cleaning contacts or replacing the
in-vacuum cable may be required.

A somewhat more thorough diagnosis may be performed with the optional
Crystal Sensor Emulator, 760-601-G1. See

section 4.3 on page 4-15

for a

discussion of its use and diagnostic capabilities.A more detailed
troubleshooting guide is shipped with the sensor. Refer to that manual for
more detailed information in some cases

.

2. Noisy signal

a. Distance from sensor to
Q-pod is longer than 40 in.
(101.6 cm).

a. Replace cables to
preserve the maximum
allowable length.

b. Mechanical/electric noise
sources located near the
Q-pod/cable .

b. Look for potential noise
sources. Reposition the
Q-pod/cables accordingly.

3. Frequency reading is
unstable or drifting

a. Temperature (of the
crystal) is changing. An
AT-cut crystal frequency may
drift as much as 10 Hz/°C.

a. Control the temperature of
the chamber. Check watering
cooling for flow and
temperature. Check source
to crystal distance (12 in. is
ideal).

b. Humidity (level on the
crystal) is changing. Moisture
being absorbed or exuded
from the crystal surface,

b. Control the humidity of the
chamber.

Table 4-1 General Troubleshooting

SYMPTOM

CAUSE

REMEDY

Advertising