4.2.2 Cool Drawer
Follow the procedure below to replace the crystal in a Next Generation Cool
Drawer™ sensor:
1
Using your thumb and index fingers, gently squeeze the sides of the retainer at
the mid section then lift it up, away from the drawer, as shown in
.
2
Hold the drawer by the handle and turn it upside down to remove the spent
crystal.
3
Install a new crystal in the drawer. Observe its orientation. The pattern
electrode should face upward as shown in
.
4
Hold the retainer by its sides. Align its orientation notch with the drawer then
gently and evenly push the retainer down until it snaps firmly into the drawer.
see
. Never push down (or pull up) on the contact spring, doing so
may permanently damage it.
5
Inspect the whole assembly. The retainer should be even and engage the
drawer at all four corners.
Figure 4-2 Cool Drawer - Removing The Crystal
Contact Spring