4.2.2 Cool Drawer

Follow the procedure below to replace the crystal in a Next Generation Cool
Drawer™ sensor:

1

Using your thumb and index fingers, gently squeeze the sides of the retainer at
the mid section then lift it up, away from the drawer, as shown in

Figure 4-2

.

2

Hold the drawer by the handle and turn it upside down to remove the spent
crystal.

3

Install a new crystal in the drawer. Observe its orientation. The pattern
electrode should face upward as shown in

Figure 4-3

.

4

Hold the retainer by its sides. Align its orientation notch with the drawer then
gently and evenly push the retainer down until it snaps firmly into the drawer.
see

Figure 4-3

. Never push down (or pull up) on the contact spring, doing so

may permanently damage it.

5

Inspect the whole assembly. The retainer should be even and engage the
drawer at all four corners.

Figure 4-2 Cool Drawer - Removing The Crystal

Contact Spring