where:

T

f

= thickness of deposited film (kÅ)

F

co

= starting frequency of the sensor crystal (Hz)

F

c

= Final frequency of the sensor crystal (Hz)

F

q

= Nominal blank frequency = 6045000 (Hz)

z = Z-ratio of deposited film material

Z

q

= Specific acoustic impedance of quartz = 8765000 (MKS units)

p = density of deposited film (g/cc)

For multiple layer deposition (for example, two layers), the Z-value used for the
second layer is determined by the relative thickness of the two layers. For most
applications the following three rules will provide reasonable accuracies:

If the thickness of layer 1 is large compared to layer 2, use material 1 Z-value
for both layers.

If the thickness of layer 1 is thin compared to layer 2, use material 2 Z-value for
both layers.

If the thickness of both layers is similar, use a value for Z-Ratio which is the
weighted average of the two Z values for deposition of layer 2 and subsequent
layers.