XTC/C - XTC/2 Operating Manual

2.7.4 Nontraditional Applications

In addition to their normal application as a deposition monitor/controller, quartz
crystal microbalances have significant utility as generalized mass sensors. This
particular instrument family is capable of measuring mass increases or
decreases on the face of the monitor crystal to an accuracy of +/- 0.617
nanograms/cm

2

(density = 1.00, z = 1.00). As always, it is imperative that the

mass be well adhered to the face of the crystal or improper readings will be
taken. It is especially important to recognize this requirement for measurements
of liquids or other non-rigid materials. INFICON’s 6MHz crystal holders have an
open area of ~0.535 cm

2

. For the highest accuracy possible, it is suggested that

the individual crystal holder be measured with a traveling microscope to
determine the exact opening area.

2.7.4.1 Etching

The instrument may be configured to display the thickness or mass removed
from the face of a crystal. It is imperative that the material be removed uniformly
over the active area of the crystal or improper readings will be taken. This
inaccuracy occurs because of radial mass sensitivity differences across the
face of the monitor crystal.

The etch mode is established by setting a configuration switch (refer to

section

2.6.2 on page 2-19

) on the back of the instrument.

The unit is operated normally, with the ZERO or START keys used to zero the
displayed thickness. The FINAL THK parameter may be programmed to
terminate the process.

2.7.4.2 Immersion in Liquids

Measurement of mass change in liquids is a relatively new field, consequently
application information is limited. The energy loss from the vibrating crystal into
the liquid environment is high, limiting the accuracy of the measurement in
some cases. The ModeLock oscillator again provides superior performance,
allowing operation in liquids of higher viscosity than an active oscillator system
would provide. The presence of bubbles on the face of the crystal as it is
immersed will drastically change the noted frequency shift and alter the
sensitivity of the technique from immersion to immersion.

NOTE: It is not recommended to use standard INFICON sensors in liquids

without modification.

2.7.4.3 Biological

The measurement of biological specimens is subject to many of the same
problems as covered in the measurement of liquids.