INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 25

Advertising
background image

tira85e1 (2013-06) PxG55x.cp

25

High Trip Point

0
1

Disable
Enable

Low Trip Point

0
1

Disable
Enable

High Trip Point Limit: High limit to trigger trip point condition if Input Value (I
0x900E, SI 0x01) is above this limit.

Object index of High Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800E1100 (High Trip Point Limit).

For ATM Mode set value to 0x800E1300 (Percentage High Trip Source).

Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x6010, SI
0x11) in [%].

Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900E,
SI 0x01) is below this limit.

Object index of Low Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800E1400.

For ATM Mode set value to 0x800E1600.

High Trip Point Limit: High limit to trigger trip point condition if Input Value (I
0x900E, SI 0x01) is above this limit.

High Trip Hysteresis: Hysteresis value for High Trip Point.

Low Trip Hysteresis: Hysteresis value for Low Trip Point.

Object index of active source of (I 0x900E SI 0x01) Input Value.

Source Index
Bit 16…31
Bit 08…15
Bit 00…07

Index
Subindex
Reserved


Is set to Combi Gauge Value 0xF6401100.

Subindex 0x01

Subindex 0x02

Subindex 0x11

Subindex 0x12

Subindex 0x13

Subindex 0x14

Subindex 0x15

Subindex 0x16

Subindex 0x17

Subindex 0x18

Subindex 0x1A