tira85e1 (2013-06) PxG55x.cp

2 Interface

Connection

For operating the Pirani Standard (PSG55x) or Pirani Capacitance Diaphragm
(PCG55x) Gauge via EtherCAT, two interface cables conforming to the EtherCAT
standard are required.
If no such cables are available, make two according to the following indications.

Ethernet Patch Cable or Crossover Cable (CAT5e quality) with FCC68 connector.


Pin

assignment:

8

1

FCC68, 8-pin,
male, soldering side

Pin Signal Description

1

TD+

Transmission Data +

2

TD-

Transmission Data -

3

RD+

Receive Data +

4

nu

not

used

5

nu

not

used

6

RD-

Receive Data -

7

nu

not

used

8

nu

not

used

Pin assignment of the D-Sub 15-pin sensor connector according to the
respective operating manual (

→  [1], [2]).

Plug the EtherCAT (and sensor) cables connector into the gauge: From the

previous device the cable connected to OUT port has to be connected to the
PxG55x <IN> port. And the cable from the PxG55x <OUT> port has to be
connected to the next device's <IN> port.

FCC68 cable

<IN> port

FCC68 cable
<OUT> port

Sensor cable


Making an EtherCAT interface
cable