The Information Data object defines the input process data.


Indicated

pressure

Real pressure

p6'
p5'

p4'
p3'

p2'
p1'

p0 p1

p2 p3

Error

Overrange

Underrange

Error

Dis

play ra

ng

e

1

2

3

4

Measurement

range

The "measurement range" is the range between minimum and maximum pressure,
where the reading of the gauge is within the specified measurement uncertainty
limits.

The display range is the complete pressure range where the gauge gives an
indication (measurement signal). The display range consists of underrange,
measurement range and overrange.

According to these definitions, 4 points are defined:

1

Lowest informational measurement value

2

Lowest precision measurement value

3

Highest precision measurement value

4

Highest informational measurement value


Index

SI

DataType

NV Access PM Name

9000 0x02

REAL x

RO Highest

Informational

Measurement Value

0x03

REAL

x

RO

Highest

Precision

Measurement Value

0x04

REAL

x

RO

Lowest

Precision

Measurement Value

Highest Informational Measurement Value: Highest value that the gauge can
measure without a specified accuracy.

Highest Precision Measurement Value: Highest valid value at which the gauge is
specified with an accuracy value. Above that value, the parameter Reading Valid is
set to invalid.

Lowest Precision Measurement Value: Lowest valid value at which the gauge is
specified with an accuracy value. Below that value, the parameter Reading Valid is
set to invalid.

4.10 PCG with ATM

Information Area
(0x9000…0x9FFF)

Highest / lowest
measurement value

Subindex 0x02

Subindex 0x03

Subindex 0x04