4 configuration trip point 2 – INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 26

Advertising
background image

26

tira85e1 (2013-06) PxG55x.cp

Calculating the High Trip Point and Low Trip Point

→ 4.4.7.

Index

SI

DataType

NV Access PM Name

800F

0x01 BOOL

x

RW

High Trip Enable

0x02 BOOL

x

RW

Low Trip Enable

0x11 REAL

x

RW

High Trip Point Limit

0x12 UDINT

x

RW

High Trip Source Index

0x13 REAL

x

RW

Percentage High Trip
Source

0x14 REAL

x

RW

Low Trip Point Limit

0x15 UDINT

x

RW

Low Trip Source Index

0x16 REAL

x

RW

Percentage Low Trip
Source

0x17 REAL

x

RW

High Trip Hysteresis

0x18 REAL

x

RW

Low Trip Hysteresis

0x1A

UDINT

x

RW

Source

Index

High Trip Point

0
1

Disable
Enable

Low Trip Point

0
1

Disable
Enable

High Trip Point limit: High limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is above this limit.

Object index of High Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800F1100 (High Trip Point Limit).

For ATM Mode set value to 0x800F1300 (Percentage High Trip Source).

Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x60010, SI
0x11) in [%].

Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is below this limit.

Object index of Low Trip Point Value source.

Bit 16…31: Index
Bit 08…15: Subindex
Bit 00…07: reserved

For a fix limit set value to 0x800F1400.

For ATM Mode set value to 0x800F1600.

High Trip Point Limit: High limit to trigger trip point condition if Input Value (I
0x900F, SI 0x01) is above this limit.

4.7.4 Configuration Trip

Point 2

Subindex 0x01

Subindex 0x02

Subindex 0x11

Subindex 0x12

Subindex 0x13

Subindex 0x14

Subindex 0x15

Subindex 0x16

Advertising
This manual is related to the following products: