18 adjust and reset, 1 command zero adjust – INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 54

Advertising
background image

54

tira85e1 (2013-06) PxG55x.cp

Measurement principle assigned to the object instance.
The most significant nibble of the parameter represents the sensor type of the first
module, the second most significant nibble of the parameter represents the sensor
type of the second Module, and so forth.

Measurement Principle (Index F940)

1
2
3
4
5

Capacitance Manometer
Piezo
Pirani
Cold Cathode
Hot Cathode

Number of Sensors (Index F940): The number of sensors implemented on the
device.

If the device consists out of several software portions this parameter should be
used.
Using standard a.b.c.d format to describe a version:

SW Version VPG PCB (Index F940)

a
b
c
d

Mayor revision
Minor revision
Development revision
Vendor specific

i.e. 1.0.1.2



Execution of this command will start a Zero Adjust operation.

Index

SI

DataType

NV Access PM Name

FB40 0x01

V_STRING

RW Command

0x02

BYTE

RO

Status

0x03

V_STRING

RO

Response

Command
Byte 0

0: Zero adjust with no offset
1: Zero adjust with zero offset
2: Zero adjust with target offset

Byte 1

Index of the Sub Sensor Instance (1)

Byte 2…5 Offset value (Data format: REAL, always 0)

Status (supported values)

0
1
2

3

255

Last command completed, no errors, no reply available
Last command completed, no errors, reply available
Last command completed, errors present, no reply
available
Last command completed, errors present, reply available
Command is executing

Subindex 0x01 (F940)

Subindex 0x02 (F940)

Subindex 0x04 (F940)

4.18 Adjust and Reset

4.18.1

Command Zero Adjust

Subindex 0x01

Subindex 0x02