1 configuration capacitance diaphragm, 2 configuration trip point 1 – INFICON SKY CDGxxxD (EtherCAT) User Manual

Page 17

Advertising
background image

tira68e1 (2013-05) CDGxxxD.cp

17

Status High Trip

0
1

High Trip not assert
High Trip assert

Status Low Trip

0
1

Low Trip not assert
Low Trip assert




Index

SI

DataType

NV Access PM Name

8001

0x12 REAL

x

RW

Offset Customer Specific

Customer-specified Offset which shall be added to the Value parameter of the
Analog Input Sensor instance.

The Offset Customer Specified parameter is a value added from a Zero Adjust
Service to the reported pressure value.

Value shall be calculated as:

Value

*)

= Reported pressure value + AZO

**)

+ OCS

***)

*)

Value (I: 0x6000, SI: 0x11)

**)

A

ccumulated Zero Offset (I: 0x9000, SI: 0x01)

***)

O

ffset Customer Specified (I: 0x8001, SI: 0x12)


The High Trip Point is calculated by:

High Trip Point = Value

*)

× Percentage

**)

*)

Value from Piezo Sensor (I: 0x6010, SI: 0x11)

**)

Percentage

High Trip Source (I: 0x800E/F, SI: 0x13)

In this case, the High Trip Point value is automatically updated by the device itself.
Any user-written value to the parameter High Trip will be refused with an abort
code.

The Low Trip Point is a fixed value:

The value defined in Low Trip Point Limit is compared with the pressure value
referenced by the Source Index parameter.

Subindex 0x01

Subindex 0x02

4.5 Configuration Area

(0x8000…0x8FFF)

4.5.1 Configuration

Capacitance Diaphragm

Subindex 0x12

4.5.2 Configuration Trip

Point 1