3 configuration trip point 2 – INFICON SKY CDGxxxD (EtherCAT) User Manual

Page 19

Advertising
background image

tira68e1 (2013-05) CDGxxxD.cp

19

Object index of active source of (I 0x900E SI 0x01) Input Value.

Source Index
Bit 16…31
Bit 08…15
Bit 00…07

Index
Subindex
Reserved


Calculating the High Trip Point and Low Trip Point

→ 4.4.7.

Index

SI

DataType

NV Access PM Name

800F

0x01 BOOL

x

RO

High Trip Enable

0x02 BOOL

x

RO

Low Trip Enable

0x11 REAL

x

RO

High Trip Point Limit

0x13 REAL

x

RW

Percentage High Trip
Source

0x14 REAL

x

RW

Low Trip Point Limit

0x17 REAL

x

RO

High Trip Hysteresis

0x18 REAL

x

RW

Low Trip Hysteresis

0x1A

UDINT

x

RO

Source

Index

High Trip Point

0
1

Disable
Enable

Low Trip Point

0
1

Disable
Enable

High Trip Point limit: High limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is above this limit.

Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x60010, SI
0x11) in [%].

Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900F,
SI 0x01) is below this limit.

High Trip Hysteresis: Hysteresis value for High Trip Point.

Low Trip Hysteresis: Hysteresis value for Low Trip Point.

Object index of active source of (I 0x900F SI 0x01) Input Value.

Source Index
Bit 16…31
Bit 08…15
Bit 00…07

Index
Subindex
Reserved


Subindex 0x1A

4.5.3 Configuration Trip

Point 2

Subindex 0x01

Subindex 0x02

Subindex 0x11

Subindex 0x13

Subindex 0x14

Subindex 0x17

Subindex 0x18

Subindex 0x1A