1 information capacitance diaphragm, 2 information common, 3 information piezo – INFICON SKY CDGxxxD (EtherCAT) User Manual

Page 21

Advertising
background image

tira68e1 (2013-05) CDGxxxD.cp

21

Index

SI

DataType

NV Access PM Name

9001 0x01

UINT

RO Sensor

Warning

0x02

UINT

RO

Sensor

Alarm

Sensor Warnings
Bit 0
Bit 1…8
Bit 9
Bit 10…13
Bit 14
Bit 15

Not at temperature
0
Electronics Warning
0
Low Warning Exception
High Warning Exception

Sensor Alarms
Bit 0…8
Bit 9
Bit 10…13
Bit 14
Bit 15

0
Electronics Failure
0
Low Alarm Exception
High Alarm Exception


Index

SI

DataType

NV Access PM Name

9010

0x01 REAL

x

RO

Accumulated Zero Offset

0x02

REAL

x

RO

Highest

Informational

Measurement Value

0x03

REAL

x

RO

Highest

Precision

Measurement Value

0x04

REAL

x

RO

Lowest

Precision

Measurement Value

Accumulated Zero Offset: An amount added prior to Gain to derive Value. Result of
Zero Adjust Command (accumulated value).

Highest Informational Measurement Value: Highest value that the gauge can
measure without a specified accuracy.

Highest Precision Measurement Value: Highest valid value at which the gauge is
specified with an accuracy value. Above that value, the parameter Reading Valid is
set to invalid.

Lowest Precision Measurement Value: Lowest valid value at which the gauge is
specified with an accuracy value. Below that value, the parameter Reading Valid is
set to invalid.

Index

SI

DataType

NV Access PM Name

9012 0x01

UINT

RO Sensor

Warning

0x02

UINT

RO

Sensor

Alarm

Sensor Warnings
Bit 0…15

0

Sensor Alarms
Bit 0…15

0

4.6.1 Information

Capacitance Diaphragm

Subindex 0x01

Subindex 0x02

4.6.2 Information

Common

Subindex 0x01

Subindex 0x02

Subindex 0x03

Subindex 0x04

4.6.3 Information

Piezo

Subindex 0x01

Subindex 0x02

Advertising